National Repository of Grey Literature 4 records found  Search took 0.01 seconds. 
Interferometric System for Length Calibration on a Metre Scale
Jedlička, Petr ; Mikel, Břetislav ; Řeřucha, Šimon ; Matěj, Z. ; Herman, O. ; Holá, Miroslava ; Jelínek, Michal ; Pavelka, Jan ; Číp, Ondřej ; Lazar, Josef
Interferometric system is designed to calibrate the measurement frame, used for periodic assessment of the inner geometry of the reactor active zone chassis. Calibration lenght is up to 3300mm.
Frequency stabilized semiconductor laser source for high-resolution interferometry
Řeřucha, Šimon ; Hucl, Václav ; Holá, Miroslava ; Čížek, Martin ; Pham, Minh Tuan ; Pravdová, Lenka ; Lazar, Josef ; Číp, Ondřej
We have assembled an experimental iodine stabilized Distributed Bragg Reflector (DBR) diode based laser system lasing at a wavelength that is in a close proximity to the wavelength of a stabilized He-Ne lasers traditionally used for metrological applications (λ=632.9 nm in vacuum). The aim was to verify whether such a system could be used as an alternative to the He-Ne laser while yielding wider optical frequency tuning range, higher output power and high frequency modulation capability. We have measured the basic characteristics of the laser source and then we have compared the performance of the laser system with that of a traditional frequency stabilized He-Ne laser with a series of experimental arrangements similar to those usually found in laser interferometry and displacement metrology applications. The results indicate that DBR diode laser system provides a good laser source for applications in dimensional (nano)metrology since it provides more output power and advanced tunability options than stabilized He-Ne lasers while maintaining fundamental requirements such as the frequency stability, coherence length and also a defined traceability.\n\n
Measurement of real-time gigacycle fatigue through real-time interferometry
Lazar, Josef ; Holá, Miroslava ; Hrabina, Jan ; Číp, Ondřej
We present an interferometric method for measurement of strain induced deformation of metal samples in experiments inducing high-cycle fatigue. The motivation is to get real-time information about the deformation of a metal sample under test, its elongation imposed by fast vibrating actuator. Fatigue of materials is a process of degradation of a material due to repeated application of forces which includes nucleation and propagation of cracks and leads to the failure of the component. It may appear in all parts of machines or structures which rotate, vibrate, are repeatedly loaded, are subjected to temperature gradients etc., so statistically, fatigue is implied in about 80% of all industrial failures.
Interferometric Displacement Measurement with Frequency Modulated Beam
Řeřucha, Šimon
In this paper, we describe a new detection method, based on the traditional homodyne detection. The focus of the new method consists in using a frequency-modulated laser source for extracting phase information with similar accuracy. Unlike the original techniques the new method does not require to use polarizing optics in the interferometer itself, and only one photodetector is sufficient for the evaluation.

Interested in being notified about new results for this query?